Thermal shock requirement specification
Operating ambient temperature from 0 (10)°С till 40°С
Relative humidity at a temperature of 23°С from 40% till 80%
Consumed power, kW, not more than 6,4
Supply voltage 380V З phase 50Hz
Maximum current, А, not more than 17
Amount of released heat, kJ/hour, not more than 17,585
Working zone dimensions (WxHxD), mm not less than 320*148x230
Working zone volume, L (liters) 11
Working zone lift capacity, kg 8
Exterior dimensions (WхHхD), mm not more than 680x1625x1050
Maximum noise, dB, not more than 60
Amount of noise at a rang of 1 m from the chamber, dB 50
Mass, kg, not more than 390
Temperature range, °С -65 -:- +200
Moving working zone:
- High temperature range zone, °С +60 -:- + 200
- Low temperature range zone, °С -65 -:- 0
- Temperature fluctuation, °С ±0,5
Heating zone:
- Temperature limit, °С + 200
Cooling zone:
- Temperature limit, °С - 80
Performance:
- Heating rate in a heating zone from an ambient temperature till +200°С, min, not more than 30
- Cooling rate in a cooling zone from an ambient temperature till - 80°С, min 90
- Temperature change time on a testing sample for two zones, min 5